Abstract

A method for the analysis of airborne (semi-)volatile polyfluorinated alkyl substances (PFASs) was optimized to avoid solvent-induced response enhancements as observed using ethyl acetate as extraction solvent (recoveries up to 300%, likely due to acetic acid as impurity). Of nine solvents tested, only the use of acetone, dichloromethane, methyl tert-butyl ether, and acetone:petroleum ether (1:1, v/v) resulted in recoveries below 100% and acetone:methyl tert-butyl ether (1:1, v/v) was chosen as suited extraction solvent. An appropriate GC capillary column and the application of mass-labelled internal standards appeared to be essential for these analyses. Instrumental limits of detection of <0.2 pg (dimethylperfluoroocatanesulfonamide, ethyl-[ 2H 5]perfluorooctanesulfonamide, methyl-[ 2H 3]perfluorooctanesulfonamide) to 8.2 pg (perfluorooctanesulfonamide) and LOQ of 0.4 pg (dimethylperfluoroocatanesulfonamide, ethyl-[ 2H 5]perfluorooctanesulfonamide, methyl-[ 2H 3]perfluorooctanesulfonamide) to 16.4 pg (perfluorooctanesulfonamide) were determined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.