Abstract
Methotrexate (MTX) is an antineoplastic drug, and due to its high toxicity, the therapeutic drug monitoring is strictly conducted in the clinical practice. The chemometric optimization and validation of a high performance liquid chromatography (HPLC) method using core–shell particles is presented for the determination of MTX in plasma during therapeutic monitoring. Experimental design and response surface methodology (RSM) were applied for the optimization of the chromatographic system and the analyte extraction step. A Poroshell 120 EC-C18 (3.0mm×75mm, 2.7μm) column was used to obtain a fast and efficient separation in a complete run time of 4min. The optimum conditions for the chromatographic system resulted in a mobile phase consisting of acetic acid/sodium acetate buffer solution (85.0mM, pH=4.00) and 11.2% of acetonitrile at a flow rate of 0.4mL/min. Selectivity, linearity, accuracy and precision were demonstrated in a range of 0.10–6.0µM of MTX. The application of the optimized method required only 150µL of patient plasma and a low consumption of solvent to provide rapid results.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.