Abstract

ABSTRACTOperational parameters for a single particle soot photometer (SP2) and a CETAC Marin-5 nebulizer were optimized for detection of particulates aerosolized from liquid samples. The sensitivity of nebulization efficiency on nebulizer input gas pressure, liquid sample flow rate, and alcohol doping of the sample were explored. The nebulization efficiency of the Marin-5 was found to be roughly independent of applied gas pressure once above a minimum pressure. The nebulization efficiency changed by ∼50% for an order of magnitude change in liquid sample flow. Doping the sample with isopropyl alcohol at a 1:1 ratio results in a ∼50% relative increase in nebulization efficiency over a broad range of liquid flows. These results should apply to all particulate materials in the size range studied. SP2 operational parameters including sheath and sample flow were explored to optimize detection of refractory black carbon (rBC) specifically via coupling to the nebulizer. The SP2 tested samples up to 5 cc s−1 with ...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.