Abstract

We used bacterial cellulose membranes as outer membranes for glucose sensors based upon amperometric detection of glucose oxidase reaction products. We found a long-term stability in 1:10 diluted blood of about 200 h. In comparison a Cuprophan® membrane was stable for only 30 h under the same conditions. Comparing the two membranes in undiluted human blood, a similar trend was found in the long-term stability: Cuprophan® was stable for only 3–4 h and the glucose sensor with the bacterial cellulose membrane was stable for more than 24 h. In addition, the measuring range of the glucose sensor could be extended up to 170 mM glucose by treating the bacterial cellulose membrane with a polyamide.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.