Abstract

A lung counter system with four 5 in. dia × 5 in. NaI cylindrical detectors, which are coupled to four lead collimators to ensure uniformity of response in space, has been developed and was described in a previous paper. The problem with this counter is the dynamic transfer of activity from the lung to other neighboring body compartments. In order to solve this problem, maximum separation between lung and neighboring compartments must be obtained without losing much of the system's sensitivity. We have accomplished this through geometric improvements in the lead collimators and with a newly proposed counting procedure. The counting procedure is based on formulating two equations, which are then solved to get the true lung activity. As a result, the interference between the measured activity in the lung and the measured activity in the stomach is rendered negligible. This decreases the uncertainty of the measured lung activity, which is of importance in lung deposition and clearance studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.