Abstract

Abstract. A new coaxial device with 7-mm-N-connector was developed providing calculable complex reflection coefficients for traceable calibration of vector network analyzers (VNA). It was specifically designed to fill the gap between 0 Hz (DC, direct current) and 250MHz, though the device was tested up to 10GHz. The frequency dependent reflection coefficient of this device can be described by a model, which is characterized by traceable measurements. It is therefore regarded as a “traceable model". The new idea of using such models for traceability has been verified, found to be valid and was used for these investigations. The DC resistance value was extracted from RF measurements up to 10 GHz by means of Genetic Algorithms (GA). The GA was used to obtain the elements of the model describing the reflection coefficient Γ of a network of SMD resistors. The DC values determined with the GA from RF measurements match the traceable value at DC within 3·10-3, which is in good agreement with measurements using reference air lines at GHz frequencies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.