Abstract

Abstract This article describes the evaluation of optimal operating modes for three existing respirable dust samplers by computer modeling of their experimentally determined sampling characteristics. The motivation behind this study is the agreement, in principle, to replace the diverse United States, European, and International conventions for respirable dust sampling with a single new convention. In assessing the sampler performance with respect to the new convention, some questions of general importance in the development of performance specifications for dust sampling instruments are discussed. The three instruments considered in this study were the U.K. SIMPEDS cyclone, the U.S. 10-mm nylon cyclone, and the MRE 113A horizontal elutriator. Although there is a considerable body of experimental data on these instruments, in no case is the complete picture available. The experimental data were used to computer model the mass that the instruments would sample from a range of ideal lognormally distributed ...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.