With the aim to achieve fast and accurate online measurement of particle size and concentration in different high-pressure gas, the application of an optical particle counter (OPC) is extended from atmospheric pressure to high-pressure conditions (more than 2MPa). But with the increase of the gas pressure, the performance of the OPC is reduced, resulting in the error of the tested particle size. In order to resolve this problem, an optimization model of optical window thickness is established for the optical sensor of the OPC based on the geometric optics theory. The effects of gas pressure, temperature, and medium on the optimized model are analyzed, which are helpful to improve the accuracy of the optical sensor. Finally, the feasibility, reliability, and accuracy of the method are verified through experiments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.