Abstract

A computational code, based on the genetic algorithm and MCNPX version 2.6 code was developed and used to investigate the effects of some important parameters of HPGe detector (such as Al cap thickness, dead-layer thickness and Ge hole size) on optimum dimensions of marinelli beaker. In addition, the effects of detector material on optimal beaker dimensions were also investigated. Finally, the optimized beaker dimensions at various beaker volumes (300, 500, 700, 1000 and 1500 cm3) were determined for some conventional Ge detectors with different crystal sizes (16 sizes). These sets of data then were used to drive mathematical formulas (obtained by best fitting to data sets).The results showed that, there is no meaningful correlation between the optimum dimensions of the beaker and each of the dead-layer thickness, Al cap thickness and the Ge-crystal hole size. On the other hand, the optimum beaker radius increases with decreasing the density of the detector material while the beaker height decreases.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.