Abstract
SYRMA-3D (SYnchrotron Radiation MAmmography 3D) aims to develop a breast CT system based on monochromatic synchrotron radiation and a single photon counting detector (PIXIRAD-8) with CdTe sensor. Due to the demanding requests on high contrast resolution and low dose, images in breast CT are particularly sensitive to small imperfections of the flat field correction applied before the CT reconstruction. Detectors based on high Z crystal sensors show inhomogeneous pixels gain, which depends on the time from the switching-on of the high voltage polarization. This effect has been studied in our CdTe detector with the purpose of develop an effective flat field correction procedure. In the PIXIRAD-8 detector, the time-dependent inhomogeneities of the flat field signal appear to be local, small and systematically reproducible, with the exception of the pixels on the sensors edges.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.