Abstract
A method was developed and validated for the analysis of benzene, toluene, ethylbenzene, xylene (BTEX), and styrene in sea water samples using gas chromatography-mass spectrometry (GC-MS) with static headspace extraction. To carefully characterize the performance of this method, several factors affecting the parameters were studied in detail, such as the heating procedure, loop filling time, equilibrium temperature, the sample dilution ratio, equilibrium, and injection times. The optimized conditions are as follows: 1)the heating procedure, 50°C for 4 min, then increased to 170°C at 15°C min-1, 2) equilibrium temperature, 60°C, 3) equilibrium time, 5 min, and 4) injection time 0.5 min. Under the optimized conditions, the linear equations were obtained in the concentration range of 10 - 50 μg L-1 with good correlation coefficient values. The limits of detection were 1.8 - 3.3 μg L -1 . The recoveries at the spiked level of 20 μg L -1 BTEX and styrene ranged from 97% - 103% with the relative standard deviations were (RSD, n=3) 1.0% - 7.6%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.