Abstract
This study aims to establish optimal scan parameters by high temporal resolution computed tomography (CT) scan for emergency patients who cannot hold their breath. First, we investigated scan parameters that can reduce the effect of motion by evaluating motion artifacts from the moving phantom scan and the temporal sensitivity profile (TPS) measurement. Second, we confirmed the standard deviation (SD) of the CT values as well as the operating time and exposure dose. As the results, plan C [rotation time: 0.275 s, detector rows: 80, pitch factor (PF): 1.100] and plan E (rotation time: 0.275 s, detector rows: 100, PF: 0.880) demonstrated high temporal resolution. The difference between the two is PF. The noise of plan C increased because PF is higher than plan E. This is also evident from the results of SD measurement. Our study demonstrates that the optimal parameters for patients who cannot hold their breath in the emergency care are plan C and plan E. In conclusion, we clarified necessary optimal scan parameters to provide clinical image that has more diagnostic information by reducing the effect of breath motion for emergency patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.