Abstract

Single distance X-ray propagation-based phase-contrast imaging is considered as a simple method compared to those requiring additional precise instruments and sophisticated algorithms to retrieve phase images. It requires, however, a modicum of conditions within the setup which include partial coherence and small pixel size at the sample position. While these conditions are usually satisfied at synchrotron light sources, they are not always satisfied within laboratory setups. In fact, these setups are limited by the size of the polychromatic source that directly influences the partial coherence of the beam, the propagation distance and the photon flux. A prior knowledge of the sample refractive index, namely the ratio of delta (δ) and beta (β) values, are also essential for the phase retrieval but this method is powerful in the presence of noise compared to absorption-based imaging. An investigation of the feasibility and the efficient applicability of this method in a commercially available X-ray microscope is conducted in this work.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.