Abstract

Abstract We experimentally generate polarization spectroscopy (error) signals corresponding to the D2-line hyperfine transitions, F g = 2 → F e = 1 , 2 , 3 of 87Rb, and F g = 3 → F e = 2 , 3 , 4 of 85Rb, and show that the strongest error signals correspond to the closed hyperfine transitions (oscillator strength ∼ 0.7 ), F g = 2 → F e = 3 and F g = 3 → F e = 4 , respectively. We make the generated error signal robust to fluctuations in external parameters by finding optimum values for the vapor cell temperature and pump-probe intensities at two different beam diameters. We further employ these optimized error signals to directly (without the need for frequency/phase modulation) lock the laser frequency—reducing the rms drift/linewidth from ∼ 10 MHz to < 500 kHz, when measured over a ∼ 60 min duration. In addition, by comparing theoretically calculated and experimentally measured error signals for the pump intensities ranging from ∼ 0.1 I sat 0 − 10 I sat 0 (where, I sat 0 ∼ 1.6 mW cm−2 is the two-level saturation intensity for the Rb D2-line transitions), we discuss the applicability and limitations of existing numerical and analytical approaches based on a full multi-level rate equation model for polarization spectroscopy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.