Abstract

In this study, the optimization of the p-type HPGe detector model has been performed to improve the accuracy of Monte Carlo simulation in calculating full energy peak efficiency. The optimized detector model was validated by comparison of simulated efficiencies with experimental efficiencies for measurement configurations of point-like sources at various distances from the detector surface and cylindrical sources on the detector surface. The obtained results showed an excellent agreement between the simulated and experimental efficiencies with relative deviations of less than 4% in the energy range from 53 to 1847 keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.