Abstract

A technique has been developed to detect and determine multi heavy metals simultaneously in a water sample. Hydride generating technique was implemented to convert the analyte which present in the water sample (liquid phase) into another form with an improved separation coefficient, called “derivative”. This process occurred without changing the original chemical structure. Derivatives were separated from the liquid phase by applying custom made gas-liquid separator (GLS), operated with oscillation. Separated species then transferred into a die-electric barrier discharge (DBD) plasma atomizer where a fragmentation of the analyte into free atoms is occurred. The generated atoms were detected by emission spectroscopy. The presented technique was applied for detection of individual and multi heavy metals simultaneously in water sample and proved useful in terms of reducing the effect of the hydrogen generated, through the process, on suppressing the atoms signal in the DBD atomizer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.