Abstract

In applications where random multi-photon events must be distinguishable from the background, detection of the signals must be based on either analog current measurement or photon counting and multi-level discrimination of single and multi-photon events. In this paper a novel method for optimizing photomultiplier (PMT) pulse discrimination levels in single- and multi-photon counting is demonstrated. This calibration method is based on detection of photon events in coincidence to short laser pulses. The procedure takes advantage of Poisson statistics of single- and mult-iphoton signals and it is applicable to automatic calibration of photon counting devices on production line. Results obtained with a channel photomultiplier (CPM) are shown. By use of three parallel discriminators and setting the discriminator levels according to the described method resulted in a linear response over wide range of random single- and multi-photon signals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.