Abstract

Abstract Rotameter is one of the most commonly used local indicating type flow measuring instrument. For remote indication and control a secondary transducer like Hall Probe sensor, LVDT etc. is incorporated with the conventional rotameter. In this paper, a modified rotameter with Hall Probe sensor is used as a measuring instrument. The output hall voltage is proportional to the flow rate of the fluid and the change in fluid density may also vary the hall voltage. So this kind of variation shows incorrect flow rate if the density of the float is not taken to a very high value compared to the density of the fluid. But the density float may affect the flow rate measurement and introduce error. In this respect firstly the variation of Hall voltage with respect to liquid density is analyzed and then the measuring system is calibrated using ANN. The ANN calculates the correction factor with respect to the change in liquid density, which results in obtaining the output close to the desired output. The simulation results show that the calibration technique is efficient.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.