A moderator of paraffin wax assembly has been demonstrated where its thickness can be optimized to thermalize fast neutrons. The assembly is used for measuring fast neutron flux of a neutron probe at different neutron energies, using BF3 (Φ1″ and 2″) and 3He(Φ0.5″) neutron detectors. The paraffin wax thickness was optimized at 6 cm for the neutron probe which contains an Am–Be neutron source. The experimental data are compared with Monte Carlo simulation results using MCNP5 version 1.4. Neutron flux comparison and neutron activation techniques are used for measuring neutron flux of the neutron probe to validate the optimum paraffin moderator thickness in the assembly. The neutron fluxes are measured at (1.17 ± 0.09) × 105 and (1.19 ± 0.1) × 105 n/s, being in agreement with the simulated values. The moderator assembly can easily be utilized for essential requirements of neutron flux measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.