Abstract
The optimum conditions for the determination of manganese at low concentrations have been developed employing differential pulse polarography. The limit of quantification of 1.04 µg/ml was achieved. The possible interference of coexisting metal ions was examined and ruled out. The method has been successfully applied for the analysis of manganese in industrial waste samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.