Abstract

This paper deals with searching of HPLC chromatographic conditions for determination and separation of Transkarbam 12 (T 12) and its two main degradation products (ω-aminocaproic acid and dodecylalcohol). T 12 is a new substance which belongs to the group of accelerators of transdermal penetration. Chromatographic separation was achieved using Separon SGX C18 analytical column (150 mm × 3 mm i.d.; 5 μm). Mobile phase contained acetonitrile and sodium acetate buffer pH 4.5 at the flow rate of 1 ml/min. Separation was carried out under the conditions of gradient elution. After the modification of the structure by derivatization reagent (3,5-dinitrobenzoyl chloride) detection at wavelength 230 nm was realized. The aim of this study was not only the optimization of the separation of derivatization reagent and derivatized T 12, Ak and D but also optimal derivatization processes for all three substances.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.