Abstract

The neutron diffraction properties of highly oriented pyrolytic graphite (HOPG) are reviewed using experimental results that have been obtained by diffraction of high-energy gamma rays, X-rays and neutrons. The interpretation of the empirical data based on diffraction theory leads to generic diagrams that display the performance of HOPG as a function of crystal thickness, mosaic spread and neutron wavelength. The analysis of the relation between the defect structure and diffraction properties demonstrates the usefulness of a detailed X-ray diffraction study to maximize the efficiency of composite neutron monochromators and analyzers. The optimization procedure is illustrated by the configuration of a double-focusing monochromator.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.