Abstract
High-frequency magnetic force microscopy (HF-MFM) is, e.g., employed for the analyses of harddisk write heads. To improve the performance of HF-MFM, we developed and analyzed cantilevers coated by (Ni, Zn) ferrites. The magnetic coatings were prepared by RF sputtering with a typical thickness of 50 nm. We systematically compared the HF-MFM images of harddisk write heads obtained by ferrite- and CoCr-coated cantilevers. The ferrite-coated cantilevers were found to produce increased HF-MFM signals up to the GHz regime, whereas the contrast in the images taken by the conventional CoCr-coated cantilevers appear to be strongly reduced at frequencies above 400 MHz. The improvement of the HF-MFM performance due to the ferrite-coated cantilevers is discussed by considering the specific imaging mechanism.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.