Abstract
ABSTRACTA detailed description is provided for the optimization of heat treatment and calibrating procedure for Idaho National Laboratory’s high temperature irradiation resistant thermocouples (HTIR-TC). Also discussed is the implication of the procedures on the overall performance of the HTIR-TC finished product; in particular, for the case of long lead sections that are typical of sensors deployed in nuclear reactors. The effect on the measurement accuracy of fluctuations of the reference temperature and localized heating along the sensor lead is also investigated. A calibration graphical user interface (GUI) and accompanying script is presented for unifying the calibration process. A fifth-order polynomial function was found to be best for fitting HTIR-TC calibration data. The performance of the HTIR-TCs subject to the optimized heat treatment and calibration procedures is shown to be robust and consistent with other TC industry standards. Electromotive force curves versus length of the TC are presented to clarify positioning of TCs in general and how its effect on the calibrated temperatures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.