Abstract
The separation of trace level actinides has been evaluated on extraction chromatography columns. Detection of the actinides was achieved through the use of an inductively coupled plasma MS (ICP-MS). The columns that we tested were prepared from a commercial TRU resin. The separation of the actinides was optimized for several parameters including particle size, column length, packing pressure, and eluent flow rate. We also examined the possibility of reducing or eliminating oxalic acid in the eluents in order to improve the performance of the mass spectrometer. We were able to separate a mixture of five actinides ((232)Th,( 238)U,( 237)Np, (239)Pu,( 243)Am) in less than 4 min. This work has application to rapid bioassay as well as for automated separations of actinide materials.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.