Abstract

A generic performance comparison strategy has been developed to evaluate the impact of mobile-phase additives (ion-pairing agent / counter ion systems), distinct stationary phases on resulting resolving power, and MS detectability of oligonucleotides and their critical impurities in gradient IP-RPLC. Stationary-phase considerations included particle type (core-shell vs. fully porous particles), particle diameter, and pore size. Separations were carried out at 60°C to optimize mass transfer (C-term). The incorporation of an active column preheater mitigated thermal mismatches, leading to narrower peaks and overcoming peak splitting. Acetonitrile as organic modifier outweighed methanol in terms of peak-capacity generation and yielded a 30% lower back pressure. Performance screening experiments were conducted varying ion-pairing agents and counter ions, while adjusting gradient span achieved an equivalent effective retention window. Hexafluoromethylisopropanol yielded superior chromatographic resolution, whereas hexafluoroisopropanol yielded significantly higher MS detection sensitivity. The 1.7 µm core-shell particle columns with 100 Å pores provided maximum resolving power for small (15–35 mers) oligonucleotides. Sub-min analysis for 15–35 polyT ladders was achieved operating a 50 mm long column at the kinetic performance limits. High-resolution separations between a 21-mer modified RNA sequence oligonucleotides and its related (shortmer and phosphodiester) impurities and complementary strand were obtained using a coupled column set-up with a total length of 450 mm.