Abstract

The influences of detection device geometry and fiber optic parameters on near infrared spectroscopy measurements were assessed using stone fruit models based on Monte Carlo simulation. The stone fruit was modeled as concentric spherical layered tissues including the skin, the flesh and the core. The choices of the detection angle, the diameter of the detection fiber, the numerical aperture, and the height of the probe were discussed. Receiving diffuse reflectance signals at detection angles in the range of 35°–50° and normalizing the detection signals by the collection area and the solid acceptance angle prior to use are suggested. Fiber probes with diameters D = 0.06 cm or 0.1 cm, NA = 0.20 or 0.30, and height h ≤ 0.8 cm are preferred. The probe deflection angle should be limited to within ±5° to guarantee measurement accuracy.