Abstract

Laminar optical tomography (LOT) is a mesoscopic tomographic imaging technique ranging between confocal microscopy and diffuse optical tomography (DOT). Fluorescence LOT (FLOT) provides depth-resolved molecular information with 100–200 μm resolution over 2–3 mm depth. In this study, we use Monte Carlo simulation and singular-value analysis (SVA) to optimize the source-detector configurations for potential enhancement of FLOT imaging performance. The effects of different design parameters, including source incidence and detector collection angles, detector number, and sampling density, are presented. The results indicate that angled incidence/detection configuration might improve the imaging resolution and depth sensitivity, especially for low-scattering medium. Increasing the number of detectors and the number of scanning steps will also result in enhanced imaging performance. We also demonstrate that the optimal imaging performance depends upon the background scattering coefficient. Our result might provide an optimization strategy for FLOT or LOT experimental setup.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.