Abstract

This study presents the advantages of the 20 microm inner diameter (id) capillary for the enantioseparation of ten basic drugs with native beta-CD as the chiral selector. The apparent binding constants of each enantiomeric pair were determined to calculate the optimum beta-CD concentration ([beta-CD]opt) and the optimization was subsequently carried out. Comparison of the 20 microm id with 50 microm id were made in terms of the results obtained in the optimization and detection limits. Applying the optimum conditions for each compound, reproducible results (RSD from 0-3; n>5) were obtained for the 20 microm id capillary. Although the sensitivity is lower in the 20 microm id capillary, the LOD determined using this capillary is still found to be acceptable for the ten basic drugs studied. Enhanced resolution and faster analysis times were the main advantages observed with the use of this capillary in enantioseparation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.