Abstract

Evaluation of radiation exposure and image quality during fluoroscopy using a new vascular X-ray system. The measurements were made on an Integris V 3000 X-ray system with MRC tube and SpectraBeam technology (Philips Medical Systems, Hamburg). Entrance dose rates were measured with phantoms for the three fluoroscopy levels (1-3) which differed with regard to beam filtering and image intensiver entrance exposure rate. We evaluated 132 diagnostic and interventional angiographic studies. The angiographic investigators were asked to start with level 1 and to change to the next fluoroscopy level only in the case of insufficient image quality. Entrance dose rate is reduced by approx. 74% at fluoroscopy level 1 and by approx. 46% at level 2 relative to level 3 which is comparable to angiographic X-ray systems without MRC tube and SpectraBeam technology. Because level 1 ensured a sufficient image quality in 92% of the diagnostic and 60% of the interventional angiographic procedures a change to higher fluoroscopy levels was not necessary. Reduction of the intensifier exposure rate and the optimization of beam filtering enabled us to reduce the radiation exposure considerably. The procedure was well accepted by the angiographic investigators due to the diagnostically sufficient image quality of the fluoroscopy level 1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.