Abstract

The conventional measurement systems for low frequency (50 - 60 Hz) high electric fields measurement currently lay on active metallic probes which can disturb the measured electric field. The present study concerns an optical device using a LiNbO<SUB>3</SUB> crystal, which is known to be electrooptic, as sensing medium. The experimental set-up uses two electrodes to create the electric field to be measured. The principle of measurement implements the electrooptical effect in the crystal without any contacting electrode and is based on the classical static method in which the polarization of light is analyzed after going through the crystal. In order to minimize thermo-optical effects, the light beam propagates along the optical axis and the electric field is applied perpendicularly to the beam. Thanks to an excellent agreement between simulation with Finite Element Method and experiment, the shape of the crystal and its protection ring are optimized. The crystal used here has a regular octagonal section (radius #1.5 mm) and a length of 8 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.