Abstract
Abstract. In this work, it is shown how a finite element method (FEM) model of a Tian–Calvet calorimeter is used to find improvements in the sensor design to increase the sensitivity of the calorimeter. By changing the layout of the basic part of the sensor, which is a low temperature co-fired ceramics (LTCC) based sensor disc, an improvement by a factor of 3 was achieved. The model was validated and the sensors were calibrated with a set of measurements that were later used to determine the melting enthalpies and melting temperatures of indium and tin samples. Melting temperatures showed a maximum deviation of 0.2 K while the enthalpy was measured with a precision better than 1 % for most samples. The values for tin deviate by less than 2 % from literature data.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.