Abstract

A radiochemical procedure is described for the fast and sensitive measurement of uranium isotopes in gaseous and liquid effluents of nuclear facilities. Equally, this procedure is suitable to measure uranium isotopes in all kinds of environmental samples. Uranium is leached from ashed sample materials with HNO3, HF, and Al(NO3)3 solution and separated from matrix elements by extraction with trioctylphosphinic oxide and backextraction with NH4F. After radiochemical cleaning by coprecipitation with LaF3 and anion exchange, uranium isotopes are electroplated on stainless steel discs from HCl/oxalate solution. The preparation is measured by alpha-spectrometry using surface barrier detectors. The detection limit for 1000 minutes of counting time is 2 mBq per sample and nuclide, the chemical yield is in the range of 50 to 80%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.