Abstract
In a search to optimize neutron long counter design for overall efficiency and flat energy response, Monte Carlo simulations were carried out for a variety of detector design parameters using the Monte Carlo N-Particle Extended code. Based on the standard long counter design by McTaggart, moderator diameter, moderator back length, and longitudinal hole diameter were sequentially varied, and the sensitivity of each parameter to the long counter response was systematically analyzed. For each design, simulations were done in the neutron energy range of 1 keV to 10 MeV. From the simulation results, it turned out that out of the three moderator parameters, the moderator diameter is most sensitive for optimizing the long counter response. As the last design parameter, the effect of the central slow-neutron counter was investigated, which showed a significant difference in the response. The investigation of each design parameter gave clear insight on its effect on the long counter response and enabled one to determine the optimum condition.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.