Abstract
A selective and sensitive analytical procedure for rapid arsenic determination by gas-diffusion flow injection analysis with amperometric detection was developed. The method is based on the arsenite reduction by NaBH(4). Derived arsine diffuses through a PTF membrane into the acceptor flow stream and is amperometrically determined on a platinum working electrode. The limit of detection (3 sigma) at room temperature was 5 microg/dm(3) of As(III). The relative standard deviation for a 1 mg/dm(3) As(III) standard was 1.96% for six repetitive injections. Arsenic(V) was determined after its prereduction with potassium iodide. Arsenic determination was not interferred with by 1 mg/dm(3) Sb(III), 5 mg/dm(3) Sn(II), 10 mg/dm(3) Se(IV), 1 mg/dm(3) As(V), 1 mg/dm(3) hydrasine, 1 mg/dm(3) Fe(II) or 0.5 mg/dm(3) Fe(III) solution. The throughput of this method was 60 analyses per hour. This method was successfully applied to arsenic determination in some power plant waste water samples.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.