Abstract

Continuous measurement of the hematocrit level in blood can potentially be performed using optical bre sensors. The Fabry Perot interferometric sensors are a promising candidate in this application. The most important step in the design of the sensor is design of the sensing interferometer. Adequate cavity length and high interference contrast are two most important requirements in this application. The design method of the sensing interferometer presented in this paper uses a Gaussian beam approximation. In order to verify the design, an optical bre Fabry Perot interferometer with adjustable cavity length was built. Its performance was tested, con rming validity of the design approach.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.