Abstract

A method for the determination of total selenium in wheat and wheat flour using graphite furnace atomic absorption spectrometry (GFAAS) with palladium/ascorbic acid as a chemical modifier was studied. The effects of nickel nitrate, palladium/ascorbic acid, and palladium/magnesium nitrate as chemical modifiers on the sensitivity in the determination of selenite, selenate and selenomethionine by GFAAS were compared. The palladium/ascorbic acid modifier was used for the determination of total selenium in wheat and wheat flour, because the oxidation states of the selenium ion are not important in the determination. The detection limit was estimated to be 1 microg L(-1) (calculated as 3sigma of the blank); the calibration curve was linear for the concentration range 5 - 50 microg L(-1) and the recovery range was 96.66 - 101.80%. The optimal ashing and atomizing temperatures were 1300 degrees C and 2250 degrees C, respectively. The proposed method was successfully applied to the determination of total selenium in wheat and wheat flour.