Abstract

This paper describes a newly developed high performance liquid chromatography/electrospray/mass spectrometry (HPLC/ES/MS) method for the determination of flunarizine (FZ) in artificial cerebrospinal fluid. The optimization for the detection of FZ in biological fluid by LC/ES/MS was investigated. The effects of solvent composition, the addition of modifier and flow rate on the detection of FZ by ES/MS were examined. The detection limit of this method (∼0.8 nM) proved to be much better than previously reported methods. Satisfactory accuracy (98.2–106.0%) of this newly developed method was obtained. The application of this method was demonstrated by analyzing FZ in rat microdialysis samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.