Abstract

With the development of the precise radiotherapy, the accuracy of radiotherapy equipment is gradually improved. The gantry, the carrier of the treatment head and various testing devices, is the most important component that determines the accuracy of the entire equipment. In this paper, the layout of the O-ring linear accelerator and the structure of the gantry are optimized to reduce the weight of the gantry by 50% and the moment of inertia by 60%. A mechanical isocenter detection method based on the laser tracker is proposed to conduct a rapid and accurate isocenter measurement for the optimized gantry. The experimental results show that the maximum deformation of the optimized gantry under load is 0.13mm during rotation and the maximum intersection distance between the gantry axis and the treatment head axis at the isocenter position is 0.21mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.