Abstract

PurposeRobotic stabilization of a therapeutic radiation beam with respect to a dynamically moving tumor target can be accomplished either by moving the radiation source, the patient, or both. As the treatment beam is on during this process, the primary goal is to minimize exposure of normal tissue to radiation as much as possible when moving the target back to the desired position. Due to the complex mechanical structure of 6 degree-of-freedom (6DoF) robots, it is not intuitive as to what 6 dimensional (6D) correction trajectory is optimal in achieving such a goal. With proportional-integrative-derivative (PID) and other controls, the potential exists that the controller may generate a trajectory that is highly curved, slow, or suboptimal in that it leads to unnecessary exposure of healthy tissue to radiation. This work investigates a novel feedback planning method that takes into account a robot’s mechanical joint structure, patient safety tolerances, and other system constraints, and performs real-time optimization to search the entire 6D trajectory space in each time cycle so it can respond with an optimal 6D correction trajectory.MethodsComputer simulations were created for two 6DoF robotic patient support systems: a Stewart-Gough platform for moving a patient’s head in frameless maskless stereotactic radiosurgery, and a linear accelerator treatment table for moving a patient in prostate cancer radiation therapy. Motion planning was formulated as an optimization problem and solved at real-time speeds using the L-BFGS algorithm. Three planning methods were investigated, moving the platform as fast as possible (platform-D), moving the target along a straight-line (target-S), and moving the target based on the fastest descent of position error (target-D). Both synthetic motion and prior recorded human motion were used as input data and output results were analyzed.ResultsFor randomly generated 6D step-like and sinusoidal synthetic input motion, target-D planning demonstrated the smallest net trajectory error in all cases. On average, optimal planning was found to have a 45% smaller target trajectory error than platform-D control, and a 44% smaller target trajectory error than target-S planning. For patient head motion compensation, only target-D planning was able to maintain a ≤0.5mm and ≤0.5deg clinical tolerance objective for 100% of the treatment time. For prostate motion, both target-S planning and target-D planning outperformed platform-D control.ConclusionsA general 6D target trajectory optimization framework for robotic patient motion compensation systems was investigated. The method was found to be flexible as it allows control over various performance requirements such as mechanical limits, velocities, acceleration, or other system control objectives.

Highlights

  • Modern radiation therapy (RT) delivery methods have evolved to the extent that technologies such as Intensity Modulated Radiation Therapy (IMRT) or Volumetric Modulated Arc Therapy (VMAT) can tightly conform radiation to the 3D shape of a target with millimeter accuracy [1, 2]

  • Computer simulations were created for two 6 degree-of-freedom (6DoF) robotic patient support systems: a Stewart-Gough platform for moving a patient’s head in frameless maskless stereotactic radiosurgery, and a linear accelerator treatment table for moving a patient in prostate cancer radiation therapy

  • Adaptive radiation therapy (ART) methods aim to address the RT motion management problem by using image guidance to locate the position of targets and adapting the radiation dose to conform to these new positions

Read more

Summary

Introduction

Modern radiation therapy (RT) delivery methods have evolved to the extent that technologies such as Intensity Modulated Radiation Therapy (IMRT) or Volumetric Modulated Arc Therapy (VMAT) can tightly conform radiation to the 3D shape of a target with millimeter (mm) accuracy [1, 2] This opens up the possibility of further radiation dose escalation to the tumor while still keeping doses low to surrounding organs-at-risk (OAR). In stereotactic radiosugery (SRS) or stereotactic body radiation therapy (SBRT) small, highly focused, and accurate radiation beams are used to escalated doses in 1-5 treatments as opposed to smaller doses over 20-30 treatments In such cases motion-related errors of even 1-2 mm can be significant. This is especially true for single-isocentermultiple-target SRS, where small rotations can lead to large positional errors for targets located

Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.