Abstract
Self-expanding Nitinol (nickel-titanium alloy) stents are tubular, often mesh like structure, which are expanded inside a diseased artery segment to restore blood flow and keep the vessel open following angioplasty. The super-elastic and shape memory properties of Nitinol reduce the risk of damage to the stent, both during delivery into the body and due to accidents while in operation. However, as Nitinol stents are subjected to a long-term cyclic pulsating load due to the heart beating (typically 4×107 cycles/year) fatigue fracture may occur. One of the major design requirements in medical implants is the device lifetime or, in engineering terms, fatigue life. The objective of this paper is to optimize the stent design by reducing the strain amplitude and mean strain over the stent, which are generated by the cyclic pulsating load. An optimization based simulation methodology was developed in order to improve the fatigue endurance of the stent. The design optimization approach is based on the Response Surface Method (RSM), which is used in conjunction with Kriging interpolation and Sequential Quadratic Programming (SQP) algorithm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.