Abstract

Summary This paper describes the optimization and validation of an analytical method for HPLC analysis of acetazolamide in tablets using Box-Behnken design. This multivariate approach enables substantial improvement of chromatographic performance using fewer experiments, without additional cost of columns or other equipment. By use of quadratic regression analysis, equations were developed describing the behavior of the response as simultaneous functions of the selected independent variables. Accordingly, the optimum conditions were determined. For quality-control samples these were use of a C18 column with acetonitrile-phosphoric acid buffer solution (0.01 m, pH 7.1) 15:85 (ν/ν) as mobile phase at a flow rate of 1.0 mL min−1. Diode-array detection was performed at 266 nm. The method was validated, in accordance with ICH guidelines, for accuracy, precision (intra-day and inter-day coefficient of variation <2.0%), selectivity, and linearity (R2 = 0.9993) over the concentration range 10–300 μg mL−1. The met...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.