Abstract
Application of polarized noble gas technology in lung functional magnetic resonance imaging (fMRI) has garnered attention for its unique advantages, such as high resolution and a lack of radiation exposure. This paper presents a 4-channel radio frequency (RF) coil design method for applications of an 1H/3He MRI system at the ultra-low field of 0.06T. For the complex model of the double-nuclear 1H/3He coil, the analytical optimization method (based on the theories of Biot-Savart law and PSO algorithm) and the electromagnetic (EM) field and radio frequency (RF) circuit co-simulation method was implemented to optimize the analysis, resulting in an effective evaluation. The simulation results demonstrated that the proposed model has the potential for imaging of the lung with the 1H/3He MRI system at an ultra-low field.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.