Abstract

Alpha-spectrometric measurements using Si detectors is the standard method for the determination of alpha emitting actinide elements. This method requires the preparation of sources for analysis which do not degrade the energy spectrum of the emitted alpha particles via sample self-absorption. A variety of methods for the electrodeposition of actinides have been reported in the literature, many of which require long deposition times and lack reproducibility. A sulfate based method has been evaluated for the preparation of these sources using chemometric analysis to optimize the method and evaluate several variables and their interactions with the goal to achieve high yield source preparation in 1 hour or less. Typical resolution for this method is 30 keV or less with recoveries approaching unity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.