Abstract
Optimization of the Finnigan GCQ ion trap mass spectrometry (ITMS) system and a clean-up procedure were carried out in order to apply high-resolution gas chromatography–tandem mass spectrometry for the analysis of polychlorinated biphenyls (PCBs) in aerosols. Six ITMS operating parameters, including isolation time, excitation voltage, excitation time, “ q” value, ion source temperature and electron energy were adjusted in order to optimize the instrument analytical performance. The adjustment of all parameters substantially increased the sensitivity of ITMS in the MS–MS mode. Changes in isolation time did not particularly affect ITMS sensitivity while ion source temperature had the strongest influence. After optimization, a limit of detection of 600 fg/μl with S/ N varying from 8 up to 91 was achieved. The application of the optimized ITMS parameters conjointly with the developed clean-up procedure resulted in method detection limits of 10–20 fg/m 3 for the determination of PCBs, in the particulate and gas phase of the atmospheric aerosol of background areas in the Eastern Mediterranean and Sweden.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.