Abstract
A series of in vitro experiments were performed to determine the efficacy of generalised phase- and amplitude-modulated sequences for low-power nonlinear microbubble contrast imaging. The microbubble agent Definity® (Dupont, Boston, MA) was exposed to sequences in which the phase and amplitude were changed from one pulse to the next. Echoes from these pulses were combined to suppress or enhance particular linear or nonlinear components. The results show that established two-pulse pulse-inversion and amplitude-modulation approaches perform similarly, providing 14 ± 1 dB of enhancement, compared with the echoes from the linear scatterer. A two-pulse combined phase and amplitude sequence achieved an additional 4 ± 1 dB of enhancement. This improvement is due to improved preservation of second and third order harmonic signals, while maintaining the suppression of the linear signals. These results were obtained at low power, below the threshold of microbubble destruction, and are applicable to real-time perfusion imaging. (E-mail: r.eckersley@imperial.ac.uk)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.