Abstract

This study sought to optimise the swallowing computed tomography (SCT) scan protocol for use with the new wide-area detector-row CT (ADCT) scanner and to estimate patient dose in terms of the organ-absorbed dose and the effective dose. The conventional ADCT (ADCTViSION) and the new ADCT (ADCTGENESIS) scanner were compared using: (1) the organ-absorbed dose and the effective dose, with a phantom study, (2) the detailed organ-absorbed doses of the neck region, using a Monte Carlo simulation and (3) a relative visual quality analysis. The effective energy differed significantly between the ADCTViSION (50 keV) and the ADCTGENESIS (57 keV). The effective doses were 2.9 and 1.9 mSv, respectively. Compared with the ADCTViSION, the absorbed dose was reduced by 34% with the ADCTGENESIS. With the ADCTGENESIS, the tube current could be reduced from 40 to 30 mA. With the optimised scan protocol, a further 25% dose reduction can be achieved.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.