Abstract
In this paper the optical parameters of Low Coherence Speckle Interferometry (LCSI) are analysed in order to optimise the technique and increase the probing depth. The contrast of the interference signal depends on the configuration of the optical setup and the optical properties of the material. Theoretical investigations and measurements for optimising the beam ratio, the coherence function and imaging parameters are presented. By optimising the interference signal the probing depth of the technique can be extended. This is demonstrated for a semi-transparent polymer material. Finally, measurements of the deformation of interfaces in a multi-layered material are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.