Abstract

To find the tube voltage that results in the highest image quality per effective dose unit for chest and pelvis radiography, respectively, using image plates. Two anthropomorphic phantoms (chest and pelvis) were imaged with several different tube voltages. The mA s settings were chosen so that the effective dose to the phantom was the same, regardless of the tube voltage, for the two examinations, respectively. The quality of the images was evaluated by six experienced radiologists using visual grading analysis. For both the chest and the pelvis examinations, the image quality increased when the tube voltage was reduced compared with the standard settings (125 and 70 kV for chest and pelvis, respectively), which were used for screen-film radiography previously. The image quality of image plate radiography can be increased by lowering the tube voltage compared with the one that was used for screen-film radiography.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.