Abstract

We compared four different skin surface landmarks, the lower margin of the right 2nd costo-sternal junction (point A); the upper margin of the right 3rd costo-sternal junction (point B); the lower margin of the right 3rd costo-sternal junction (point C); and a point 5 cm below the manubrio-sternal junction (point D), in 20 cancer patients undergoing insertion of permanent central venous catheters whose tips were placed near the superior vena cava - right atrium (SVC-RA) junction under transoesophageal echocardiography guidance. The landmark was satisfactory if it was located within 1 cm of the SVC-RA junction. Points C and D were closer to the SVC-RA junction than points A and B (p < 0.0001). However, point C had the highest incidence (C: 70%, A: 0%, B: 20%, D: 30%, p < 0.0001) of being within 1 cm of the SVC-RA junction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.